Exploitation of metallic glass as new brazing filler for Ti-based biomedical alloy was attempted. Ti-6Al-7Nb was used as a brazed material,
and candidates of bulk metallic glass brazing filler were Cu60Hf25Ti15, Mg65Cu25Gd10, Zr55Cu30Al10Ni5 and Pd40Cu30P20Ni10. Convergence
infrared-ray brazing was conducted for brazing Ti-6Al-7Nb/metallic glass in Ar atmosphere. After brazing, hardness measurement, X-ray
tomography, cross-sectional observation, artificial saliva immersion test and tensile test were performed to evaluate brazability, mechanical
property and corrosion resistance of the obtained brazing joints.
The results of brazing using these metallic glass fillers show that all the metallic glasses were brazable to Ti-6Al-7Nb except for
Mg65Cu25Gd10. Mg65Cu25Gd10, Cu60Hf25Ti15 and their joints collapsed rapidly during immersion test. Zr55Cu30Al10Ni5 joint was the best in
terms of degradation resistance; however, tensile strength was inferior to the conventional one. Pd40Cu30Ni10P20 filler and Zr55Cu30Al10Ni5 filler
and their joints did not show any collapse or tarnish during the immersion test. Pd40Cu30Ni10P20 joint showed the excellent properties in terms of
both corrosion resistance and tensile strength, which were superior to a joint brazed using Ti-15Cu-25Ni conventional filler. X-ray tomograph
indicates that fracture tends to occur in the vicinity of the brazing interface after tensile test.
The brazed metallic glass fillers were fully crystallized, excluding Pd40Cu30Ni10P20 filler. Pd40Cu30Ni10P20 brazed filler contained mapleleaf
like primary dendrite, peritectoid and a few microns interfacial reaction layer in glassy matrix. The results indicated that Pd40Cu30Ni10P20 is
promising brazing filler for dental or biomaterial devices.