Novel use of a Battery-Powered Two-Electrode Emitter Cell for Electrospray Mass Spectrometry...

by Vilmos Kertesz, Gary J Van Berkel
Publication Type
Journal
Journal Name
Journal of the American Society for Mass Spectrometry
Publication Date
Page Numbers
953 to 961
Volume
17
Issue
7

A battery-powered, controlled-current, two-electrode electrochemical cell containing a porous flow-through working electrode with high surface area and multiple auxiliary electrodes with small total surface area was incorporated into the electrospray emitter circuit to control the electrochemical reactions of analytes in the electrospray emitter. This cell system provided the ability to control the extent of analyte oxidation in positive ion mode in the electrospray
emitter by simply setting the magnitude and polarity of the current at the working electrode. In addition, this cell provided the ability to effectively reduce analytes in positive ion mode and oxidize analytes in negative ion mode. The small size, economics, and ease of use of such a battery-powered controlled-current emitter cell was demonstrated by powering a single resistor and switch circuit with a small-size, 3 V watch battery, all of which might be incorporated on the emitter cell. (J Am Soc Mass Spectrom 2006, 17, 953-961) � 2006 American Society for Mass Spectrometry