Abstract
Single crystal inelastic neutron scattering measurements are often performed using a sample environment for controlling sample temperature. One difficulty associated with this is establishing appropriate thermal coupling from the sample to the temperature controlled portion of the sample environment. This is usually accomplished via a sample can which thermally couples the sample environment to the sample can and the sample can to the sample via an exchange gas. Unfortunately, this can will contribute additional background signal to ones measurement. We present here the design of an ultrathin aluminum sample can based upon established technology for producing aluminum beverage cans. This design minimizes parasitic sample can scattering. Neutron scattering measurements comparing a machined sample can to our beverage can design clearly indicate a large reduction in scattering intensity and texture when using the ultrathin sample can design. We also examine the possibility of using standard commercial beverage cans as sample cans.