Wireless Monitoring of Liver Hemodynamics In Vivo...

by Tony Akl, Mark Wilson, Milton N Ericson, Ethan D Farquhar, Gerard Cote
Publication Type
Journal
Journal Name
Proceedings of the National Academy of Sciences
Publication Date
Page Numbers
1 to 11
Volume
1
Issue
1

Liver transplants have their highest failure rate in the first two weeks following surgery. Currently, there are no devices for continuous, real-time monitoring of the graft. In this work, we present a continuous perfusion and oxygen consumption monitor based on photoplethysmography. The sensor is battery operated and communicates wirelessly with a data acquisition computer which provides the possibility of implantation provided sufficient miniaturization. In two in vivo porcine studies, the sensor tracked perfusion changes in hepatic tissue during vascular occlusions with a root mean square error (RMSE) of 0.125 mL/min/g of tissue. We show the possibility of using the pulsatile wave to measure the arterial oxygen saturation similar to pulse oximetry. This signal is used as a feedback to extract the venous oxygen saturation from the DC levels. Arterial and venous oxygen saturation changes were measured with an RMSE of 2.19 and 1.39% respectively when no vascular occlusions were induced. This error increased to 2.82 and 3.83% when vascular occlusions were induced during hypoxia. These errors are similar to the resolution of the oximetry catheter used as a reference. This work is the first realization of a wireless perfusion and oxygenation sensor for continuous monitoring of hepatic perfusion and oxygenation changes.