Skip to main content
SHARE
Publication

Characterization and Optimization of a Spectral Window for Direct Gaseous Uranium Hexafluoride Enrichment Assay Using Laser-Induced Breakdown Spectroscopy

by George Chan, Leigh R Martin, Richard Russo
Publication Type
Journal
Journal Name
Applied Spectroscopy
Publication Date
Page Numbers
819 to 834
Volume
77
Issue
8

Through a systematic scanning of 235U and 238U emission lines between 280 nm and 745 nm, the optimal emission line for direct gaseous uranium hexafluoride (UF6) enrichment assay using laser-induced breakdown spectroscopy (LIBS) was found. Screening for spectral features that are potentially useful for U isotopic analysis was gauged from the magnitude of the 235U–238U isotopic shift and the signal-to-background ratio of the emission line through a parameter termed ΔSBR 235U–238U. The ΔSBR spectrum shows peaks at wavelength positions where there are strong lines with significant 235U–238U shifts. The screening identified 13 spectral-window candidates, which were down selected based on their overall accuracy in predicting the 235U enrichment of three UF6 samples of natural (0.720 atom% 235U) and low-enriched (4.675 atom% and 9.157 atom% 235U) grades. The U(I) 646.498 nm emission line, with a determined 235U–238U isotopic shift of −17.7 pm, was found to be the optimal spectral window for direct UF6 enrichment assay. The root mean square error for enrichment assays on the three natural and low-enriched UF6 samples, with each sample measured in six replicates, was 0.31% in absolute 235U content. Each measurement comprised LIBS signals accumulated from 3000 laser shots. The analytical bias and precision were better than 0.5% and 0.3%, respectively, in absolute [235U/(235U + 238U)] ratios. Specific for the two low-enriched UF6 samples, the relative standard deviations from six replicated measurements were around 2%.