Skip to main content
SHARE
Publication

An integrated portable Raman sensor with nanofabricated gold bowtie array substrates for energetics detection ...

Publication Type
Journal
Journal Name
Analyst, The
Publication Date
Page Numbers
1697 to 1702
Volume
8
Issue
136

An integrated field-portable surface enhaned Raman scattering (SERS) sensing system has been developed and evaluated for quantitative analysis of energetics such as perchlorate (ClO4-) and trinitrotoluene (TNT) at environmentally-relevant concentrations and conditions. The detection system consists of a portable Raman spectrometer equipped with an optical fiber probe that is coupled with novel elevated gold bowtie nanostructural arrays as a sensitive and reproducible SERS substrate. Using the standard addition technique, we show that ClO4- and TNT can be quantified at concentrations as low as 0.66 mg/L (or ~6.6 µM) and 0.20 mg/L (~0.9 µM), respectively, in groundwater samples collected from selected military sites. This research represents the first step toward the development of a field SERS sensor which may permit rapid, in-situ screening and analysis for various applications including national security, chemical, biological and environmental detection.