Skip to main content
SHARE
Publication

Measurements of Scatter Peaks in 137Cs and 60Co Sources...

by Leticia S Pibida, David L West
Publication Type
Journal
Journal Name
Journal of Research of the National Institute of Standards and Technology
Publication Date
Page Number
123012
Volume
123

Results from tests of radiation detection instruments with radionuclide identification capabilities will depend on the sources used for the tests. Radionuclide identification detectors are designed to measure photons and provide an identification of the source being measured. High-resolution spectra need to be acquired to determine all the observable peaks in the source spectra before testing these types of instruments. These peaks may be due to impurities and/or scatter peaks in the sources. This paper discusses the issues encountered with the response of a radioisotope identification device due to scatter peaks in one type of source used for testing. In addition, it provides spectra for different source types and source constructions to compare the differences in scatter, allowing for a better source type selection for instrument testing.