Skip to main content
SHARE
Publication

Performance Evaluation of the Scent Transfer Unit (STU) for Organic Compound Collection and Release...

Publication Type
Journal
Journal Name
Journal of Forensic Sciences
Publication Date
Page Numbers
780 to 789
Volume
51
Issue
4

The Scent Transfer UnitTM (STU-100) is a portable vacuum that uses airflow through a sterile gauze pad to capture a volatiles profile over evidentiary items for subsequent canine presentation to assist law enforcement personnel. This device was evaluated to determine its ability to trap and release organic compounds at ambient temperature under controlled laboratory conditions. Gas chromatography-mass spectrometry (GC-MS) analyses using a five-component volatiles mixture in methanol injected directly into a capture pad indicated that compound release could be detected initially and three days after time of collection. Additionally, fifteen compounds of a 39-component toxic organics gaseous mixture (10-1,000 ppbv) were trapped, released, and detected in the headspace of a volatiles capture pad after being exposed to this mixture using the STU-100 with analysis via GC-MS. Component release efficiencies at ambient temperature varied with the analyte; however, typical values of approximately 10 percent were obtained. Desorption at elevated temperatures of reported human odor/scent chemicals and colognes trapped by the STU-100 pads was measured and indicated that the STU-100 has a significant trapping efficiency at ambient temperature. Multivariate statistical analysis of subsequent mass spectral patterns was also performed.